Description

The Mbiotech One-Step Kit has been formulated for highly reproducible first-strand cDNA synthesis and subsequent real-time PCR in a single tube. A combination of the latest advances in buffer chemistry together with a reverse transcriptase and hot-start DNA polymerase system, ensures that Mbiotech One-Step Kit produces fast, highly-specific and ultra-sensitive one-step RT-qPCR.
Mbiotech  Hi-ROX One-Step Kit consists of a 2X Hi-ROX One-step mix(ROX for optional use), as well as separate reverse transcriptase and RNase Inhibitor.
Mbiotech Lo-ROX One-Step Kit consists of a 2X Lo-ROX One-step mix(ROX for optional use), as well as separate reverse transcriptase and RNase Inhibitor.
Mbiotech No-ROX One-Step Kit consists of a 2X No-ROX One-step mix, as well as separate reverse transcriptase and RNase Inhibitor.

Kit Components

 
Reagent
100 x 20ul
500 x 20ul
One-Step mix(2x)
1 x 1 ml
5 x 1 ml
RNase Inhibitor
1 x 40 ul
1 x 200 ul
Reverse Transcriptase
1 x 20 ul
1 x 100 ul
DEPC-H2O
1 x 1.8 ml
2 x 1.8 ml

Kit compatibility

The 2x Mbiotech Hi ROX One-Step Kit has been optimized for use in Green-based real-time RT-PCR on the real-time instruments listed in the following compatibility table, each of these instruments having the capacity to analyze the real-time PCR data with the passive reference signal either on or off. The kit is also compatible with several instruments that do not require the use of ROX, such as the Qiagen (Corbett) Rotor-GeneTM 6000, the Bio-Rad CFX96 or the Roche LightCycler¢ç 480.
Manufacturer
Model
ABI
7000, 7300, 7700, 7900, 7900HT, StepOneTM,  StepOneTM Plus
 
 
 
 
 
The 2x Mbiotech Lo-ROX One-Step Kit has been optimized for use in Green-based real-time RT-PCR on the real-time instruments listed in the following compatibility table, each of these instruments having the capacity to analyze the real-time PCR data with the passive reference signal either on or off. The kit is also compatible with several instruments that do not require the use of ROX, such as the Qiagen (Corbett) Rotor-GeneTM 6000, the Bio-Rad CFX96 or the Roche LightCycler¢ç 480.
 
Manufacturer
Model
ABI (Invitrogen)
7500, 7500 FAST
Stratagene (Agilent)
Mx4000TM, Mx3000PTM, Mx3005PTM
 
 
 
 
 
 
The 2x Mbiotech No-ROX One-Step Kit is premixed with Green I dye and is compatible with real-time instruments that do not need a passive reference signal for normalized for use on the real-time instruments listed in the following compatibility table:
 
Manufacturer
Model
Bio-Rad
OpticonTM, Opticon2TM, MiniOpticon, Chromo4TM, CFX96, CFX384
Cepheid
SmartCyclerTM
Qiagen (Corbett)
Rotor-GeneTM 3000 & 6000
Eppendorf
Realplex
Roche
LightCycler¢ç 480
Techne
Quantica¢ç

 

 

 

 

 

 

  

Storage and Stability

The Mbiotech One-Step Kit is shipped on Dry/Blue Ice. All kit components should be stored at -20¡É upon receipt. Excessive freeze/thawing is not recommended. When stored under optimum conditions, the reagents are stable for a minimum of 6 months from date of purchase.

Quality Control

The Mbiotech One-Step Kit and its components are extensively tested for activity, processivity, efficiency, heat activation, sensitivity, absence of nuclease contamination and absence of nucleic acid contamination.

Safety Precautions

Harmful if swallowed. Irritating to eyes, respiratory system and skin. Please refer to the material safety data sheet for further information.
 

General considerations

 
To help prevent any carry-over DNA contamination we recommend that separate areas be maintained for PCR set-up, PCR amplification and any post-PCR gel analysis. It is essential that any amplified PCR product should not be opened in the PCR set-up area.
Primers: the sequence and concentration of primer as well as the amplicon length can be critical for specific amplification, yield and overall efficiency of any real-time PCR. We strongly recommend taking the following into consideration when designing and running your PCR reaction:
 
 
¡¤ Use primer-design software, such as Scitools (http://eu.idtdna.com/scitools/scitools.aspx), Primer3 or visual OMPTM (http://frodo.wi.mit.edu/primer3/ and DNA Software, Inc; http://dnasoftware.com/ respectively). Primers should have a melting temperature (Tm) of approximately 60¡ÆC
¡¤ Optimal amplicon length should be 50-150bp
¡¤ A final primer concentration of 250nM is suitable for most PCR conditions, however to determine the optimal concentration we recommend a primer titration in the range of 0.1–1uM
¡¤ Use equimolar primer concentrations
¡¤ When amplifying from cDNA use gene-specific primers. If possible use intron-spanning primers to avoid amplification from genomic DNA.
 
 
Template: It is important that the DNA template is suitable for use in PCR in terms of purity and concentration.  Also, the template needs to be devoid of any contaminating PCR inhibitors (e.g.EDTA). The recommended amount of template for PCR is dependent upon the type of DNA used. The following should be considered when using genomic DNA and cDNA templates:
 
¡¤ Genomic DNA: use up to 1g of complex (e.g. eukaryotic) genomic DNA in a single PCR. We recommend using the Mbiotech Genomic DNA Mini Kit for high yield and purity from both prokaryotic and eukaryotic sources.
 
¡¤ cDNA: the optimal amount of cDNA to use in a single PCR is dependent upon the copy number of the target gene. We suggest using 100ng cDNA per reaction, however it may be necessary to vary this amount. To perform a two-step RT-PCR, we recommend using the Mbiotech cDNA Synthesis Kit for reverse transcription of the purified RNA. For high yield and purity of RNA, use the Mbiotech Total RNA Mini Kit.
 
MgCl2: The MgCl2 concentration in the 1x reaction mix is 3mM, which is optimal for SensiTaq in the majority of real-time PCR conditions. If necessary, we suggest titrating MgCl2 to a maximum of 5mM.
 
PCR controls: It is important to detect the presence of contaminating DNA that may affect the reliability of the data. Always include a no template control (NTC), replacing the template with PCR-grade water. When performing a two-step RT-PCR, set-up a no RT control as the NTC for the PCR.
 
 
 
 

Procedure

Reaction Mix Composition: Prepare a master mix. The volumes given below are based on a standard 20¥ìl final reaction mix and can be scaled accordingly.
Reaction Mix Composition: Prepare an RT-PCR master mix. The volumes given below are based on a standard 20¥ìl final reaction mix and can be scaled accordingly.
 
Reagent
Volume
Final concentration
2x SYBR One-Step Mix
10ul
1x
10uM Forward Primer
0.8ul
400nM
10uM Reverse Primer
0.8ul
400nM
Reverse Transcriptase
0.2ul
-
RNase Inhibitor
0.4ul
-
DEPC-H2O
Up to 16ul
Template 4ul   
    20 ul final volume 
Suggested RT-PCR conditions: The following RT-PCR conditions are suitable for Mbiotech SYBR No-ROX One-Step Kit with a majority of amplicons and real-time PCR instruments. However, the cycling conditions can be varied to suit different machine-specific protocols.
Mbiotech SYBR No-ROX One-Step Kit is compatible with either three-step or two-step cycling:

3- step cycling

 
Cycles
Temperature
Time
Notes
1
45¡É
10min
Reverse transcription
1
95¡É
2min
Polymerase activation
95¡É
5s
Denaturation
40
60¡É
10s
Annealing
72¡É
5s
Extension(acquire at end of step)

 2-step cycling

 
Cycles
Temperature
Time
Notes
1
45¡É
10min
Reverse transcription
1
95¡É
2min
Polymerase activation
95¡É
15s
Denaturation
40
60¡É
15s
Extension(acquire at end of step)
 
 